why we use polystyrene film for FTIR calibration?

0
Most people are only using FTIR for identification, so 
wavelength accuracy is most important.

A polystyrene film is an easy way to calibrate the entire 
wavelength range of the instrument. The film produces 
sharp absorbance bands at known wavelengths (or 
frequencies) which can be used to verify that the 
instrument's spectrum is accurate.

And also additionaly other than polystyrene we can use any 
other reference material which is having known sharp 
absorbence bands at different wavenumbers.

concept is calibrating the instrument against known 
material and known results. 